第四章 气相色谱
Chapter 4:Gas Chromatography

Gas chromatography employ inert gas,commonly Nitrogen,to carry the volatile substances through the column.There are many kinds of stationary phase of GC.We switch the column for different task,but we seldom change the carrier gas. 
气相色谱应用惰性的气体,通常是氮气,来携带挥发性物质通过色谱柱。气相色谱的固定相种类繁多。我们为了不同的任务换用不同的柱子,但很少改变载气。

The carrier gas is often store in a steel cylinder with pressure and flow regulators.The gas used in GC must be dry,so it have to pass a drying tube first.The speed of the flow will be measured by a flow meter before the column. The sample is injected by a syringe through a septum at the inject port.The gas wash the sample into the column which  is placed in a thermostatic chamber.When the analytes in the sample are carried out by the gas,they will detect by a online detector.A recorder or a computer will receive the result signal and make a graph,thus we get the chromatogram.
载气常常装在带压力和流速调节器的不锈钢瓶里。用于气相的气体必须是干燥的,所以气体必须先经过一根干燥管。流速在柱前用流速计测定。样品用进样针,在进样口通过隔膜进样。气体把样品冲入放在恒温相的柱子。当样品中的被分析物被气体带出来的时候,它们会被一个在线的检测器检测到。一个记录仪或一台电脑会接受结果信号并进行绘图,这样就得到了色谱图。

We commonly evaluate the column effciency by the number of theoretical plates(N).It’s calculated by the following equation:N=5.54*(Tr/W)^2,Tr is the retension time,which can be measured by the distance from scratch to the peak maximum point.The W here means the peak width at half peak height.
我们一般通过理论塔板数(N)来评价柱子的效能。它通过下面的公式计算:N=5.54*(Tr/W)^2,Tr是保留时间,可以通过测量色谱图上峰最高点到起点的距离来得到。W这里指峰高度一半处的峰宽度。

The resolution(R) is calculaed like this:R=2(Tr2-Tr1)/(W1+W2), here Tr1,Tr2 are the retention times, W1 and W2 are the width of two adjacent peaks.
分辨率(R)是这样计算的:R=2(Tr2-Tr1)/(W1+W2),Tr1、Tr2是保留时间,W1和W2是两个相邻峰的峰宽。

The va Deemter equation is widely used to describe the relation between HETP(height equalent to a theoretical plate) and velocity(u).
HETP=A+B/u+Cu  HETP=L/N,L is the length of the column.A,B,C are constants respectively related to the eddying flow,mass transfer,and longitudinal diffusion.
描述HETP(理论塔板高度)和流速间关系用的最多的是范氏方程。
HETP=A+B/u+Cu,HETP=L/N,L是色谱柱的长度。A、B、C是分别与涡流、传质、纵向扩散有关的常数。

The detector mostly used in GC is FID(Flame Ionization Detector).TCD(Thermal Conductivity Detector) is seldom used now.And the ECD(Electron Capture Detector) gains its popularity for pesticide residue quantification.
气相常用氢火焰离子检测器。热导检测器现在很少用了。电子捕获检测器由于农药残留定量而变得流行起来。

以下摘录一些文章和书籍中例句以供参考:
1.Obviously,complete sepatation of narrow peaks can be obtained with the peak centers closer together than if the peaks are spread.[from Handbook of Chromatography]
显然,往峰中心靠拢的比峰扩散的色谱峰更容易达到完全分离。

2.The interactions between the solute and the sorbent leading to dissimilar partition coefficients are hydrogen bonds,Debye forces,Van der Waals’ forces and specific chemical interactions.[from Handbook of Chromatography]
造成溶质和吸附剂之间不同分配系数的作用有氢键、德拜力、范德华力和专属的化学作用。

3.The optimum carrier-gas flow rate is determined by plotting HETP vs u and choosing a value at or slightly above the minimum in the curve.[from Handbook of Chromatography]
最佳载气流速是通过理论塔板高度对速度作图并选择曲线的最小值或略高的值来确定。

4.It is generally desirable to ascertain the suitability and effectiveness of the operating system when employing chromatographic methods such as pressurized liquid chromatography or gas chromatography.[from U.S. Pharmacopoeia]
当应用色谱方法如加压液相或气相时,最好先确定一下系统适应性。
5.Reactivity of the support is reduced by treatment with a silanizing reagent prior to coating with liquid phase.[from U.S. Pharmacopoeia]
担体的反应性的减低是通过在涂布固定液时先用硅烷化试剂处理来达到。
6.Unless otherwise specified, the carrier gas is nitrogen; the chromatographic column is made of stainless steel or glass; the support is acid washed and silanized diatomaceous earth or porous polymer beads of 0.18—0.25 mm, 0.15—0.18 mm or 0.12—0.15 mm in diameter; the temperature of the injection port is usually 30—50℃ higher than that of the column; the amount of solution injected is several tenth of μl to several μl; a flame ionization detector is used, its temperature is equal to or higher than that of the column, but never below 100℃ so as to avoid the condensation of moisture.[from The Pharmacopoeia of the People’s Repulic of China(Eglish Edition 1988)]
除另有规定外,载气通常是氮气;色谱柱是用不锈钢或玻璃制造的;担体是酸洗并硅烷化了的硅藻土或0.18—0.25 mm, 0.15—0.18 mm 和 0.12—0.15 mm直径的多孔性聚合物圆珠;进样口温度一般是比柱温高30~50℃;进样量是零点几微升到几微升;使用氢火焰离子检测器,它的温度是等于或高于柱温,但不得低于100℃以防湿气冷凝。

信息来源:
中国色谱网
 
免责声明:
1、本文系本网编辑转载,转载目的在于传递更多信息,并不代表本网赞同其观点和对其真实性负责。
2、如涉及作品内容、版权和其它问题,请在30日内与本网联系,我们将在第一时间删除内容或提供稿费!
 
进入论坛】【 】【打印该页】 

相关新闻
·请正确使用法定计量单位--每日二题
·test
·terminology of chromatography_HPLC Chromatography
·国内、外主要分析化学及其相关学术期刊
·term of chromatography
·terminology of chromatography_Gas Chromatography
·Paper Chromatography And Thin Layer Chromatography
·有关ampoxicillin的译文
·《色谱术语》(Terminology of Chromatography)系列文章的第一篇
·求救
 
  资讯搜索  
关 键 字:
 
  热门推荐  
·先睹为快!《中国药典》2005
·转载:分析化学、药剂学、天
·色谱基础,值得参考哦
·色谱常见问题问答
·《色谱通讯》第三、四期电子
·色谱通讯2005第一期
·茄 尼 醇
·糖类分析中文
·HPLC的两个使用问题
·Agilent公司毛细管柱简介
·色谱通讯第二期
·国内、外主要分析化学及其相
·样品制备和HPLC、LC/MS方法开
·色谱扫盲班系列讲座,胖丁丁
·色谱经验问与答系列
更多>> 

关于我们 | 广告服务 | 商铺指南 | 法律声明 | 友情链接 | 联系我们 | 网站地图 | 意见反馈
服务热线:0571-28180298 传真:0571-28905256 意见与建议 点击这里给我发消息 点击这里给我发消息 新浪微博  腾讯微博
中国色谱网 版权所有©1999-2013